{"id":6823,"date":"2016-07-27T11:02:03","date_gmt":"2016-07-27T09:02:03","guid":{"rendered":"https:\/\/blog.wika.com\/en\/?p=6823"},"modified":"2022-11-07T10:43:00","modified_gmt":"2022-11-07T09:43:00","slug":"comparative-calibration-vs-fixed-point-calibration","status":"publish","type":"post","link":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/","title":{"rendered":"Comparative calibration vs. fixed point calibration"},"content":{"rendered":"<p><strong>In temperature calibration technology, there are two methods of calibration &#8211; the comparative calibration and the fixed point calibration. The method of calibration to be used will be defined through the demands on the accuracy and the type of test instrument. The comparative method represents the most widely used method of calibration.<\/strong><\/p>\n<p>So how do the comparative calibration and fixed point calibration methods differ? Using the example of calibrating a resistance thermometer, this blog article explains both methods and thus tries to convey to the reader a better understanding.<\/p>\n<h2>The comparative calibration<\/h2>\n<p>As the name already implies, this type of calibration is carried out by comparing the resistance thermometer being tested with a higher-quality reference thermometer. The higher-quality reference should thus exhibit an accuracy of three or four times higher than the thermometer being tested. Both thermometers are placed into a single homogenous temperature source. When placing the thermometers, care should be taken to ensure a small distance between them and that the sensitive points (meaning the measuring points) are at the same height. As a temperature source, liquid baths or dry-well calibrators are generally used.<\/p>\n<h2>The fixed point calibration<\/h2>\n<p>The fixed point calibration, in contrast to the comparative calibration, is carried out against the defined triple, freezing and melting points from the International Temperature Scale of 1990 (ITS-90). The fixed points define temperatures at which the physical state of substances alter. The most important fixed point of the ITS-90 is the triple point of water. At a temperature of 0.010 \u00b0C the physical state of high-purity water is constantly changing between steam, ice and liquid water. The quantities of the physical states always remain in proportion. Generally, only national laboratories or laboratories qualifying for very small measurement uncertainties work with these fixed-point cells, as these are costly and specialised only for this application. Highly stable PRTs (platinum resistance thermometers), which will be used as reference thermometers, are calibrated using the fixed point method. The reference sensor is inserted into a closed, or also open fixed-point cell, which holds the temperature constant over a long period.<\/p>\n<p><strong>Example: fixed points from -189 to 660 \u00b0C in accordance with ITS-90:<\/strong><\/p>\n<table>\n<tbody>\n<tr>\n<td><strong>Fixed point<\/strong><\/td>\n<td><strong>Temperature<\/strong><\/td>\n<\/tr>\n<tr>\n<td>Triple point of argon<\/td>\n<td>-189,3442 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Triple point of mercury<\/td>\n<td>-38,8344 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Triple point of water<\/td>\n<td>0,01 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Melting point of gallium<\/td>\n<td>29,7646 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Freezing point of indium<\/td>\n<td>156,5985 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Freezing point of tin<\/td>\n<td>\u00a0231,928 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Freezing point of zinc<\/td>\n<td>\u00a0419,527 \u00b0C<\/td>\n<\/tr>\n<tr>\n<td>Freezing point of aluminium<\/td>\n<td>\u00a0660,323 \u00b0C<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p><strong><br \/>\nNote<br \/>\n<\/strong>Further information on our <a href=\"https:\/\/www.wika.com\/en-en\/service.WIKA\" target=\"_blank\" rel=\"noopener external\">Services \u201cTemperature calibration\u201d<\/a>\u00a0can be found on the WIKA website.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>In temperature calibration technology, there are two methods of calibration &#8211; the comparative calibration and the fixed point calibration. The method of calibration to be used will be defined through the demands on the accuracy and the type of test instrument. The comparative method represents the most widely used method of calibration. So how do [&hellip;]<\/p>\n","protected":false},"author":58,"featured_media":6824,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"footnotes":""},"categories":[5],"tags":[56,282,304],"class_list":["post-6823","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-knowhow","tag-measuring-principles","tag-sensor","tag-thermometer"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v27.2 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Comparative calibration vs. fixed point calibration<\/title>\n<meta name=\"description\" content=\"This blog article explains both methods in temperature calibration technology: the comparative calibration and the fixed point calibration\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\" \/>\n<meta property=\"og:locale\" content=\"en_GB\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Comparative calibration vs. fixed point calibration\" \/>\n<meta property=\"og:description\" content=\"This blog article explains both methods in temperature calibration technology: the comparative calibration and the fixed point calibration\" \/>\n<meta property=\"og:url\" content=\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\" \/>\n<meta property=\"og:site_name\" content=\"WIKA blog\" \/>\n<meta property=\"og:image\" content=\"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"836\" \/>\n\t<meta property=\"og:image:height\" content=\"471\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Fabian Ehing\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Fabian Ehing\" \/>\n\t<meta name=\"twitter:label2\" content=\"Estimated reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"2 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\/\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#article\",\"isPartOf\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\"},\"author\":{\"name\":\"Fabian Ehing\",\"@id\":\"https:\/\/blog.wika.com\/en\/#\/schema\/person\/83783712309975704dfcbc9184b5d0eb\"},\"headline\":\"Comparative calibration vs. fixed point calibration\",\"datePublished\":\"2016-07-27T09:02:03+00:00\",\"dateModified\":\"2022-11-07T09:43:00+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\"},\"wordCount\":421,\"commentCount\":1,\"image\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg\",\"keywords\":[\"Measuring principles\",\"Sensor\",\"Thermometer\"],\"articleSection\":[\"Know-how\"],\"inLanguage\":\"en-GB\"},{\"@type\":\"WebPage\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\",\"url\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\",\"name\":\"Comparative calibration vs. fixed point calibration\",\"isPartOf\":{\"@id\":\"https:\/\/blog.wika.com\/en\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage\"},\"image\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg\",\"datePublished\":\"2016-07-27T09:02:03+00:00\",\"dateModified\":\"2022-11-07T09:43:00+00:00\",\"author\":{\"@id\":\"https:\/\/blog.wika.com\/en\/#\/schema\/person\/83783712309975704dfcbc9184b5d0eb\"},\"description\":\"This blog article explains both methods in temperature calibration technology: the comparative calibration and the fixed point calibration\",\"breadcrumb\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#breadcrumb\"},\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage\",\"url\":\"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg\",\"contentUrl\":\"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg\",\"width\":836,\"height\":471,\"caption\":\"Comparison: calibration - fixed point calibration\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\/\/blog.wika.com\/en\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Comparative calibration vs. fixed point calibration\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\/\/blog.wika.com\/en\/#website\",\"url\":\"https:\/\/blog.wika.com\/en\/\",\"name\":\"WIKA blog\",\"description\":\"Our blog offers many interesting articles on pressure, temperature, level, force, flow measurement calibration technology and provides insights into the WIKA Group.\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\/\/blog.wika.com\/en\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-GB\"},{\"@type\":\"Person\",\"@id\":\"https:\/\/blog.wika.com\/en\/#\/schema\/person\/83783712309975704dfcbc9184b5d0eb\",\"name\":\"Fabian Ehing\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\/\/blog.wika.com\/en\/files\/2015\/03\/default_avatar-140x140.png\",\"url\":\"https:\/\/blog.wika.com\/en\/files\/2015\/03\/default_avatar-140x140.png\",\"contentUrl\":\"https:\/\/blog.wika.com\/en\/files\/2015\/03\/default_avatar-140x140.png\",\"caption\":\"Fabian Ehing\"},\"url\":\"https:\/\/blog.wika.com\/en\/author\/ehingf\/\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Comparative calibration vs. fixed point calibration","description":"This blog article explains both methods in temperature calibration technology: the comparative calibration and the fixed point calibration","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/","og_locale":"en_GB","og_type":"article","og_title":"Comparative calibration vs. fixed point calibration","og_description":"This blog article explains both methods in temperature calibration technology: the comparative calibration and the fixed point calibration","og_url":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/","og_site_name":"WIKA blog","og_image":[{"width":836,"height":471,"url":"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg","type":"image\/jpeg"}],"author":"Fabian Ehing","twitter_misc":{"Written by":"Fabian Ehing","Estimated reading time":"2 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#article","isPartOf":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/"},"author":{"name":"Fabian Ehing","@id":"https:\/\/blog.wika.com\/en\/#\/schema\/person\/83783712309975704dfcbc9184b5d0eb"},"headline":"Comparative calibration vs. fixed point calibration","datePublished":"2016-07-27T09:02:03+00:00","dateModified":"2022-11-07T09:43:00+00:00","mainEntityOfPage":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/"},"wordCount":421,"commentCount":1,"image":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage"},"thumbnailUrl":"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg","keywords":["Measuring principles","Sensor","Thermometer"],"articleSection":["Know-how"],"inLanguage":"en-GB"},{"@type":"WebPage","@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/","url":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/","name":"Comparative calibration vs. fixed point calibration","isPartOf":{"@id":"https:\/\/blog.wika.com\/en\/#website"},"primaryImageOfPage":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage"},"image":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage"},"thumbnailUrl":"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg","datePublished":"2016-07-27T09:02:03+00:00","dateModified":"2022-11-07T09:43:00+00:00","author":{"@id":"https:\/\/blog.wika.com\/en\/#\/schema\/person\/83783712309975704dfcbc9184b5d0eb"},"description":"This blog article explains both methods in temperature calibration technology: the comparative calibration and the fixed point calibration","breadcrumb":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#breadcrumb"},"inLanguage":"en-GB","potentialAction":[{"@type":"ReadAction","target":["https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/"]}]},{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#primaryimage","url":"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg","contentUrl":"https:\/\/blog.wika.com\/en\/files\/2015\/09\/vergleich_fixpunkt_838_471px_v1.jpg","width":836,"height":471,"caption":"Comparison: calibration - fixed point calibration"},{"@type":"BreadcrumbList","@id":"https:\/\/blog.wika.com\/en\/knowhow\/comparative-calibration-vs-fixed-point-calibration\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/blog.wika.com\/en\/"},{"@type":"ListItem","position":2,"name":"Comparative calibration vs. fixed point calibration"}]},{"@type":"WebSite","@id":"https:\/\/blog.wika.com\/en\/#website","url":"https:\/\/blog.wika.com\/en\/","name":"WIKA blog","description":"Our blog offers many interesting articles on pressure, temperature, level, force, flow measurement calibration technology and provides insights into the WIKA Group.","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/blog.wika.com\/en\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-GB"},{"@type":"Person","@id":"https:\/\/blog.wika.com\/en\/#\/schema\/person\/83783712309975704dfcbc9184b5d0eb","name":"Fabian Ehing","image":{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/blog.wika.com\/en\/files\/2015\/03\/default_avatar-140x140.png","url":"https:\/\/blog.wika.com\/en\/files\/2015\/03\/default_avatar-140x140.png","contentUrl":"https:\/\/blog.wika.com\/en\/files\/2015\/03\/default_avatar-140x140.png","caption":"Fabian Ehing"},"url":"https:\/\/blog.wika.com\/en\/author\/ehingf\/"}]}},"_links":{"self":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts\/6823","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/users\/58"}],"replies":[{"embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/comments?post=6823"}],"version-history":[{"count":11,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts\/6823\/revisions"}],"predecessor-version":[{"id":12381,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts\/6823\/revisions\/12381"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/media\/6824"}],"wp:attachment":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/media?parent=6823"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/categories?post=6823"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/tags?post=6823"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}