{"id":7215,"date":"2016-10-14T09:35:36","date_gmt":"2016-10-14T07:35:36","guid":{"rendered":"https:\/\/blog.wika.com\/en\/?p=7215"},"modified":"2021-12-02T09:03:53","modified_gmt":"2021-12-02T08:03:53","slug":"selection-criteria-for-mechanical-thermometers-response-time-2","status":"publish","type":"post","link":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/","title":{"rendered":"Selection criteria for mechanical Thermometers (1): Response time"},"content":{"rendered":"<p><strong>Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers &#8211; the decision on the type of thermometer alone depends on several factors. One of the most important selection criteria is the response time of the measuring Instrument.<\/strong><\/p>\n<p>When it comes to speed, there is no way around gas-actuated thermometers. As a rule of thumb, these thermometers take about 30 seconds for a temperature adjustment from 10\u00b0C to 90\u00b0C. A bimetal thermometer would display the same final value after 90 seconds. Due to their short response times, gas-actuated thermometers are particularly suitable for integration into switching requirements.<\/p>\n<p>The difference in speed is explained by the different functional principles of the two thermometer types &#8211; unlike the bimetal thermometer (which uses the principle of different expansion coefficients), the measuring system of a gas-actuated thermometer is filled with helium under pressure. Any change in temperature causes an immediate change in internal pressure in the stem. The pressure deforms the measuring tube and the deflection is transferred to the pointer via a movement.<\/p>\n\n<div data-fullscreen=\"disabled\" data-mode=\"normal\" data-oembed=\"1\" data-provider=\"youtube\" data-reset-after-played id=\"arve-youtube-vum1dy9b6l8\" style=\"max-width:400px;\" class=\"arve alignleft\">\n\t<div class=\"arve-inner\">\n\t\t<div style=\"aspect-ratio:4\/3\" class=\"arve-embed arve-embed--has-aspect-ratio\">\n\t\t\t<div class=\"arve-ar\" style=\"padding-top:75.000000%\"><\/div>\n\t\t\t<iframe allow=\"accelerometer &apos;none&apos;;autoplay &apos;none&apos;;bluetooth &apos;none&apos;;browsing-topics &apos;none&apos;;camera &apos;none&apos;;clipboard-read &apos;none&apos;;clipboard-write;display-capture &apos;none&apos;;encrypted-media &apos;none&apos;;gamepad &apos;none&apos;;geolocation &apos;none&apos;;gyroscope &apos;none&apos;;hid &apos;none&apos;;identity-credentials-get &apos;none&apos;;idle-detection &apos;none&apos;;keyboard-map &apos;none&apos;;local-fonts;magnetometer &apos;none&apos;;microphone &apos;none&apos;;midi &apos;none&apos;;otp-credentials &apos;none&apos;;payment &apos;none&apos;;picture-in-picture;publickey-credentials-create &apos;none&apos;;publickey-credentials-get &apos;none&apos;;screen-wake-lock &apos;none&apos;;serial &apos;none&apos;;summarizer &apos;none&apos;;sync-xhr;usb &apos;none&apos;;web-share;window-management &apos;none&apos;;xr-spatial-tracking &apos;none&apos;;\" allowfullscreen=\"\" class=\"arve-iframe fitvidsignore\" credentialless data-arve=\"arve-youtube-vum1dy9b6l8\" data-lenis-prevent=\"\" data-src-no-ap=\"https:\/\/www.youtube-nocookie.com\/embed\/vuM1DY9b6l8?feature=oembed&amp;iv_load_policy=3&amp;modestbranding=1&amp;rel=0&amp;autohide=1&amp;playsinline=0&amp;autoplay=0&amp;enablejsapi=1\" frameborder=\"0\" height=\"300\" loading=\"lazy\" name=\"\" referrerpolicy=\"strict-origin-when-cross-origin\" sandbox=\"allow-scripts allow-same-origin allow-presentation allow-popups allow-popups-to-escape-sandbox\" scrolling=\"no\" src=\"https:\/\/www.youtube-nocookie.com\/embed\/vuM1DY9b6l8?feature=oembed&#038;iv_load_policy=3&#038;modestbranding=1&#038;rel=0&#038;autohide=1&#038;playsinline=0&#038;autoplay=0&#038;enablejsapi=1\" title=\"Response time measuring\" width=\"400\"><\/iframe>\n\t\t\t\n\t\t<\/div>\n\t\t\n\t<\/div>\n\t\n\t\n\t<script type=\"application\/ld+json\">{\"@context\":\"http:\\\/\\\/schema.org\\\/\",\"@id\":\"https:\\\/\\\/blog.wika.com\\\/en\\\/knowhow\\\/selection-criteria-for-mechanical-thermometers-response-time-2\\\/#arve-youtube-vum1dy9b6l8\",\"type\":\"VideoObject\",\"embedURL\":\"https:\\\/\\\/www.youtube-nocookie.com\\\/embed\\\/vuM1DY9b6l8?feature=oembed&iv_load_policy=3&modestbranding=1&rel=0&autohide=1&playsinline=0&autoplay=0&enablejsapi=1\",\"name\":\"Response time measuring\",\"thumbnailUrl\":\"https:\\\/\\\/i.ytimg.com\\\/vi\\\/vuM1DY9b6l8\\\/hqdefault.jpg\",\"uploadDate\":\"2021-07-20T08:58:47Z\",\"author\":\"WIKA Group\",\"description\":\"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type of thermometer alone depends on several factors. One of the most important selection criteria is the response time of the measuring Instrument. W\"}<\/script>\n\t\n<div class=\"arve-error\"><p><small><abbr title=\"Advanced Responsive Video Embedder\">ARVE<\/abbr> error: maxwidth: <code>838px<\/code> is not valid<\/small><\/p><\/div>\n<\/div>\n<h2>Thermowells affect the response time<\/h2>\n<p>As with other measuring instruments, in critical processes, mechanical thermometers must also be combined with a thermowell. It should be considered that such a sleeve will also have an influence, to some degree significant, on the response time of the thermometer. To what extent, depends on the material and wall thickness of the thermowell.<\/p>\n<p><strong>Note<br \/>\n<\/strong>Further details on mechanical thermometers can be found on the WIKA website under the headings\u00a0<a href=\"https:\/\/www.wika.com\/en-en\/dial_thermometers.WIKA\" rel=\"external\" target=\"_blank\">Dial thermometers<\/a>\u00a0and <a href=\"https:\/\/www.wika.com\/en-en\/thermometers_with_switch_contacts.WIKA\" rel=\"external\" target=\"_blank\">Thermometers with switch contacts<\/a>.<\/p>\n<p><strong>See also our articles<br \/>\n<\/strong><a href=\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-dial-thermometers-2-process-temperature\/\">Selection criteria for mechanical thermometers (2): process temperature<\/a><br \/>\n<a href=\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-dial-thermometers-3-capillary\/\">Selection criteria for dial thermometers (3): Capillary<\/a><\/p>\n<p>&nbsp;<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers &#8211; the decision on the type of thermometer alone depends on several factors. One of the most important selection criteria is the response time of the measuring Instrument. When it comes to speed, there is no way around [&hellip;]<\/p>\n","protected":false},"author":114,"featured_media":7216,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"footnotes":""},"categories":[5],"tags":[304,90],"class_list":["post-7215","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-knowhow","tag-thermometer","tag-response-time"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v27.2 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Selection criteria for mechanical Thermometers (1) - WIKA blog<\/title>\n<meta name=\"description\" content=\"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\" \/>\n<meta property=\"og:locale\" content=\"en_GB\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Selection criteria for mechanical Thermometers (1) - WIKA blog\" \/>\n<meta property=\"og:description\" content=\"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type\" \/>\n<meta property=\"og:url\" content=\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\" \/>\n<meta property=\"og:site_name\" content=\"WIKA blog\" \/>\n<meta property=\"og:image\" content=\"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"838\" \/>\n\t<meta property=\"og:image:height\" content=\"471\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"Christoph Strebel\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Christoph Strebel\" \/>\n\t<meta name=\"twitter:label2\" content=\"Estimated reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"2 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\/\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#article\",\"isPartOf\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\"},\"author\":{\"name\":\"Christoph Strebel\",\"@id\":\"https:\/\/blog.wika.com\/en\/#\/schema\/person\/f0e3c5a262124a47ee6d0ee60c9f9e0c\"},\"headline\":\"Selection criteria for mechanical Thermometers (1): Response time\",\"datePublished\":\"2016-10-14T07:35:36+00:00\",\"dateModified\":\"2021-12-02T08:03:53+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\"},\"wordCount\":292,\"commentCount\":0,\"image\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg\",\"keywords\":[\"Thermometer\",\"Time response\"],\"articleSection\":[\"Know-how\"],\"inLanguage\":\"en-GB\"},{\"@type\":\"WebPage\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\",\"url\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\",\"name\":\"Selection criteria for mechanical Thermometers (1) - WIKA blog\",\"isPartOf\":{\"@id\":\"https:\/\/blog.wika.com\/en\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage\"},\"image\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg\",\"datePublished\":\"2016-10-14T07:35:36+00:00\",\"dateModified\":\"2021-12-02T08:03:53+00:00\",\"author\":{\"@id\":\"https:\/\/blog.wika.com\/en\/#\/schema\/person\/f0e3c5a262124a47ee6d0ee60c9f9e0c\"},\"description\":\"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type\",\"breadcrumb\":{\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#breadcrumb\"},\"inLanguage\":\"en-GB\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage\",\"url\":\"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg\",\"contentUrl\":\"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg\",\"width\":838,\"height\":471,\"caption\":\"response time of thermometers\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\/\/blog.wika.com\/en\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Selection criteria for mechanical Thermometers (1): Response time\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\/\/blog.wika.com\/en\/#website\",\"url\":\"https:\/\/blog.wika.com\/en\/\",\"name\":\"WIKA blog\",\"description\":\"Our blog offers many interesting articles on pressure, temperature, level, force, flow measurement calibration technology and provides insights into the WIKA Group.\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\/\/blog.wika.com\/en\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-GB\"},{\"@type\":\"Person\",\"@id\":\"https:\/\/blog.wika.com\/en\/#\/schema\/person\/f0e3c5a262124a47ee6d0ee60c9f9e0c\",\"name\":\"Christoph Strebel\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-GB\",\"@id\":\"https:\/\/blog.wika.com\/en\/files\/2025\/10\/strebelchristoph_wiz1007-2591-10er-120x120.jpg\",\"url\":\"https:\/\/blog.wika.com\/en\/files\/2025\/10\/strebelchristoph_wiz1007-2591-10er-120x120.jpg\",\"contentUrl\":\"https:\/\/blog.wika.com\/en\/files\/2025\/10\/strebelchristoph_wiz1007-2591-10er-120x120.jpg\",\"caption\":\"Christoph Strebel\"},\"description\":\"In my role as a product manager, I am responsible for the care and the optimal strategic orientation of the product portfolio for mechanical process thermometers. Through the WIKA blog, I would like to share my experience and also my detailed technical knowledge with interested readers.\",\"url\":\"https:\/\/blog.wika.com\/en\/author\/strebec\/\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Selection criteria for mechanical Thermometers (1) - WIKA blog","description":"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/","og_locale":"en_GB","og_type":"article","og_title":"Selection criteria for mechanical Thermometers (1) - WIKA blog","og_description":"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type","og_url":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/","og_site_name":"WIKA blog","og_image":[{"width":838,"height":471,"url":"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg","type":"image\/jpeg"}],"author":"Christoph Strebel","twitter_misc":{"Written by":"Christoph Strebel","Estimated reading time":"2 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#article","isPartOf":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/"},"author":{"name":"Christoph Strebel","@id":"https:\/\/blog.wika.com\/en\/#\/schema\/person\/f0e3c5a262124a47ee6d0ee60c9f9e0c"},"headline":"Selection criteria for mechanical Thermometers (1): Response time","datePublished":"2016-10-14T07:35:36+00:00","dateModified":"2021-12-02T08:03:53+00:00","mainEntityOfPage":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/"},"wordCount":292,"commentCount":0,"image":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage"},"thumbnailUrl":"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg","keywords":["Thermometer","Time response"],"articleSection":["Know-how"],"inLanguage":"en-GB"},{"@type":"WebPage","@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/","url":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/","name":"Selection criteria for mechanical Thermometers (1) - WIKA blog","isPartOf":{"@id":"https:\/\/blog.wika.com\/en\/#website"},"primaryImageOfPage":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage"},"image":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage"},"thumbnailUrl":"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg","datePublished":"2016-10-14T07:35:36+00:00","dateModified":"2021-12-02T08:03:53+00:00","author":{"@id":"https:\/\/blog.wika.com\/en\/#\/schema\/person\/f0e3c5a262124a47ee6d0ee60c9f9e0c"},"description":"Selecting suitable mechanical thermometers for a process is not a quick task. Bimetal thermometers or gas-actuated thermometers - the decision on the type","breadcrumb":{"@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#breadcrumb"},"inLanguage":"en-GB","potentialAction":[{"@type":"ReadAction","target":["https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/"]}]},{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#primaryimage","url":"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg","contentUrl":"https:\/\/blog.wika.com\/en\/files\/2016\/05\/stoppuhr_neu.jpg","width":838,"height":471,"caption":"response time of thermometers"},{"@type":"BreadcrumbList","@id":"https:\/\/blog.wika.com\/en\/knowhow\/selection-criteria-for-mechanical-thermometers-response-time-2\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/blog.wika.com\/en\/"},{"@type":"ListItem","position":2,"name":"Selection criteria for mechanical Thermometers (1): Response time"}]},{"@type":"WebSite","@id":"https:\/\/blog.wika.com\/en\/#website","url":"https:\/\/blog.wika.com\/en\/","name":"WIKA blog","description":"Our blog offers many interesting articles on pressure, temperature, level, force, flow measurement calibration technology and provides insights into the WIKA Group.","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/blog.wika.com\/en\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-GB"},{"@type":"Person","@id":"https:\/\/blog.wika.com\/en\/#\/schema\/person\/f0e3c5a262124a47ee6d0ee60c9f9e0c","name":"Christoph Strebel","image":{"@type":"ImageObject","inLanguage":"en-GB","@id":"https:\/\/blog.wika.com\/en\/files\/2025\/10\/strebelchristoph_wiz1007-2591-10er-120x120.jpg","url":"https:\/\/blog.wika.com\/en\/files\/2025\/10\/strebelchristoph_wiz1007-2591-10er-120x120.jpg","contentUrl":"https:\/\/blog.wika.com\/en\/files\/2025\/10\/strebelchristoph_wiz1007-2591-10er-120x120.jpg","caption":"Christoph Strebel"},"description":"In my role as a product manager, I am responsible for the care and the optimal strategic orientation of the product portfolio for mechanical process thermometers. Through the WIKA blog, I would like to share my experience and also my detailed technical knowledge with interested readers.","url":"https:\/\/blog.wika.com\/en\/author\/strebec\/"}]}},"_links":{"self":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts\/7215","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/users\/114"}],"replies":[{"embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/comments?post=7215"}],"version-history":[{"count":12,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts\/7215\/revisions"}],"predecessor-version":[{"id":12375,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/posts\/7215\/revisions\/12375"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/media\/7216"}],"wp:attachment":[{"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/media?parent=7215"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/categories?post=7215"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/blog.wika.com\/en\/wp-json\/wp\/v2\/tags?post=7215"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}